High Purity Gas Analysis
Trace Gas Analysis Facility
The NPL Trace Gas Analysis Facility is capable of providing traceable measurements of the key contaminant species found in high-purity gases.
An overview of the instruments that make up the facility and their capabilities can be found in the table below:
| Trace contaminant | Instrument | Matrix gases | Detection limit (nmol/mol) |
| Water | Tiger Optics Lasertrace | Most inerts, passive gases and some toxic and corrosive gases | Sub-nmol/mol |
| Oxygen | Delta-F Nanotrace II | All inerts, H2, light HCs and halocarbons | Sub-nmol/mol |
| Carbon dioxide | Peak Performer 1 (FID + methaniser) | N2, Ar, He, O2, H2 and air | 1-2 |
| Methane | 1-2 | ||
| Non-methane hydrocarbons | 3-5 | ||
| Hydrogen | Ametek Ta5000R | N2, Ar, He, O2, propylene, ethylene and air | 1-2 |
| Carbon monoxide | 1-2 | ||
| Sulphur dioxide, nitrogen monoxide | UV analysers | N2 and air | 1 |
| Volatile organic components (VOCs) | Cryo-focus GC | N2 and air | 100 pmol/mol |
| Carbon monoxide, carbon dioxide, ethane, argon, oxygen, nitrogen, methane and hydrogen | GC-DID | N2, CH4, Ar, O2, He and air | 10 |
Measurement traceability is achieved through instrument calibrations using NPL Primary Reference Gas Mixtures. The NPL trace gas analysis facility is also available for instrument calibration and performance assessment work.