- When converting between pressure units consideration should be given to the number of significant figures to use, bearing in mind that many of the underlying conversion factors are not themselves exact and cannot be made so.
- Probably not. The figures should be interpreted with caution, especially if an instrument has not been independently calibrated for some time.
- NPL offers a cost-effective re-certification service after the initial certification period has expired.
- In short the answer is no. What you do need to take into account is the uncertainty of measurement.
- Density is defined as mass per unit volume; it has the SI unit kg·m-3 and is an absolute quantity. Specific gravity is the ratio of a material's density with that of water at 4 °C and is therefore a relative quantity with no units.
- MSF is the three-letter call sign used to designate the UK’s 60 kHz standard-frequency and time signal.
- A force measurement system is made up of a transducer and associated instrumentation.
- To calculate a pressure value using a liquid column - for example a mercury barometer - or a pressure balance it is necessary to know the gravitational acceleration at the location of the instrument. It can be determined by measurement on site, calculation or interpolation of measured values.
All FAQs
- No, they are not. The internationally recognised SI unit for pressure is the pascal, abbreviated to Pa, and this is the unit realised by the primary measurement standards in the world's national metrology institutes to provide traceability for pressure measurements.
- The International Prototype Kilogram is not perfectly stable (its mass changes with time), the amount it changes cannot be known perfectly (there is no 'perfect' reference against which to judge it) and the values of the national copies cannot be monitored at the highest level of accuracy without being compared directly with it.
- In some situations.
- The most accurate barometers are indeed the mercury primary barometers used at national measurement institutes. Most barometers, though, are secondary instruments rather than primary ones and when considering these it is not correct to say that those based on a mercury column are invariably more accurate than those that are based on an alternative principle.
- Yes there are - some are listed here.
- The time at which summer time begins and ends is given in the relevant EU Directive and UK Statutory Instrument as 1 a.m. Greenwich Mean Time (GMT).
- Up to a point yes, but unless a weight is of suitable design and material and in appropriate condition it will not be possible to give it a meaningful calibration and it would certainly be a waste of money.
- Give the weight a general inspection to check its construction, surface finish and the suitability of its magnetic properties.
- A number of factors need to be taken into account when considering sources for calibrating radiation thermometers.
- Yes - Saturated (or unsaturated) salt solutions, and certain other chemicals, can be used to generate an environment of a particular relative humidity in an enclosed space.