Troubleshooting in radiation monitoring

This module highlights problems that can be encountered in interpretation of the results and in the monitoring instruments.
A. Points to consider if interpreting the measurement of dose rate
1. Close approach to sources
As a general rule when measuring dose rate from a point source, the detector should not be placed closer than three times the detector dimensions because the radiation field is no longer uniform. A closer approach will lead to an increasing underestimate of the dose rate. Measurements can be made at greater distances and corrected for the inverse square law.
For a small area β emitting source it may be necessary to measure at a closer distance in order to detect it, but the user should be aware that there can be gross underestimates because the dose rate in contact with the source could be 100 times greater than the measured instrument indication when using a typical ion chamber instrument that has a large detector volume. An extremity dosemeter placed in close proximity to the source may provide more reliable measurements.
2. Dead Time
Consult the instruction manual to confirm whether dead time correction is required.
Most commercial instruments for the measurement of dose rate will have built in dead time corrections. However, there are some instruments in the field that do not automatically compensate for dead time, guidance on how to correct an instrument reading to account for losses due to dead time is provided.
3. Pulsed sources generating narrow pulses with durations less than 30 μs
Measuring pulsed radiation sources is a specialist task.
The use of pulse measuring equipment is not recommended in pulsed radiation fields, e.g. from linacs. The pulse from the machine is commonly only a few µs long, which means the instrument can only record one count, no matter how intense the dose rate. It is safer to use instrumentation that measures the intensity of each pulse rather than simply detecting it or not, e.g. ion chamber instruments or passive dosemeters.
However, ionisation chambers do not work well at low dose rates and passive dosemeters cannot be used to search for shielding weaknesses. To use a pulse measuring instrument expert advice should be sought to interpret the readings.
Good practice online modules
- Practical radiation monitoring
- Practical radiation monitoring units
- Radiation monitoring strategy
- Choice of radiation instrument
- Radiation monitoring techniques
- Estimating surface radioactivity
- Radiation instrument calibration
- Uncertainties in radiation measurement
- Troubleshooting in radiation monitoring


