Consider the following characteristics of an instrument

Gives an instant indication of radiation or contamination level allowing the surveyor to concentrate on guiding the probe, rather than looking at the display.
2. SensitivityThe instrument must be capable of making meaningful measurements at the lowest decision level as determined by the expected background count rate and the expected response to the radiation of interest.
For operational use, the time constant should not exceed about 4 seconds. This allows the surveyor to pause, wait about 12 seconds and then take an eye average over the following 10 seconds or so.
4. Statistical fluctuation
This has to be balanced against the time constant.
- Short time constant will give a quick response but a high level of fluctuation. This makes it difficult to record an eye average.
- Long time constant will give a relatively fixed reading, but this demands that the surveyor waits a long time between readings and eye averages over a long period.
For any monitoring process, 3 counts per second represents about the minimum level where both an acceptable time constant and an acceptable level of statistical fluctuation can be obtained.
A simple rule in the measurement of dose rate is to try to work no closer than 3 times the maximum detector dimension from the source (due to the inverse square law).
Any detector will average over its volume. If a large detector is exposed to a narrow radiation beam, the indication will be much lower than the dose rate in the beam. The best that can usually be done is to choose a suitable detector and then define the averaging area as its area. This approach favours the use of GM tubes.
7. Fail to danger
ONLY USE EQUIPMENT WITH ‘FAIL TO DANGER’ PROTECTION
Some instruments may inappropriately display zero when exposed to high dose-rate fields; this could have catastrophic consequences.
It is essential that any instrument type/instrument, has been tested up to the maximum intensity of radiation it could reasonably encounter to ensure that it continues to function correctly. Failure to danger at high dose rates is unusual in any but the oldest European equipment, but can be a problem with USA and former Soviet Union produced GM instruments in particular.
Good practice online modules
- Practical radiation monitoring
- Practical radiation monitoring units
- Radiation monitoring strategy
- Choice of radiation instrument
- Radiation monitoring techniques
- Estimating surface radioactivity
- Radiation instrument calibration
- Uncertainties in radiation measurement
- Troubleshooting in radiation monitoring
