ANAMET – Measurement Comparison No 941
Description
This exercise compared complex scattering coefficient measurements of two-port items fitted with GPC-3.5 connectors in 50 ohm coaxial line. Measurements were made over the frequency range 1 GHz to 26.5 GHz in 1.5 GHz steps. The two―port items were attenuators with nominal attenuations of 3 dB, 6 dB, 10 dB and 70 dB. Eight organisations participated in the exercise, which took place during 1994.
The exercise produced some very interesting results. For example, the results for the phase of the S11 values occasionally contained values which were nominally 180° different from the majority of values.
Reference
N M Ridler and J C Medley, "A comparison of complex scattering coefficient measurements in 50 ohm coaxial line to 26.5 GHz", NPL Report DES 138, June 1995