ARMUG Frequently Asked Questions
NPL, in conjunction with ARMUG members, has compiled the following list of questions and answers on the measurement of radioactivity in air which we hope users will find useful. Members' feedback on the content would be gratefully received, including any suggestions for additional points.
Q1A1 |
What factors should be considered when selecting a monitor?At the outset, the objective of the monitoring, and the environment in which it will take place, (especially with regard to the presence on radon isotopes) must be clearly understood. Check any available Type Test data for instruments under consideration – they must indicate that the instrument can perform adequately for your application. If the monitoring conditions are particularly demanding (for example, if concentrations of radon isotopes or solid particulates are likely to be elevated), a field trial is recommended. Also, check that the instrument is acceptable from the point of view of maintenance - spare parts must be readily available and the manufacturers or distributors must be adequately trained and able to provide back-up. Also, consider speaking to other users who have experiences with the same instrument – have they found it performs adequately? |
Q2A2 |
What are the pros and cons of different detector types?Solid state α/β-in-air monitors have the advantages that they can provide information on particle energies and enable the user to carry out α/β separation and α-spectrometry, but they are expensive, and their immunity to radiofrequencies is poor. Geiger-Müller tubes are inexpensive and simple to operate, and can be used for the assay of β-emitters, but background count-rates are too high for measuring α-emitters. α/β discrimination is not possible with these instruments, neither can they provide particle energy information. |
Q3A3 |
Where should a safe alarm limit be set to avoid false alarms?Limits of 5 – 8 DAC.h for α-emitters, and 1 DAC.h for β-emitters, are suggested. Ensure that the likely Activity Median Aerodynamic Diameter (AMAD), and the ambient radon levels, are taken into account. |
Q4A4 |
When should monitoring be carried out?Monitoring is necessary:
|
Q5A5 |
Where should the sampling head be positioned?This depends on the purpose of the measurement. When monitoring for containment failure, there should be at least one monitor positioned as close to the anticipated failure point as possible (and downstream of the air-flow from that point) and one monitor positioned remote but downstream from the point. When carrying out static sampling, monitors should be deployed across the whole work area and in positions where they are most likely to detect activity. When monitoring for dose control purposes, sampling should be done in operator’s breathing zone using a 'swan neck' or similar sampling head. |
Q6A6 |
When should alarms be set, and when should retrospective measurements be done?Alarms should be set for the higher risk operations, or when occupancy levels in the area are high. Static monitors provide assurance when working at environmental levels, and are also useful for accident consequence investigations. |
Q7A7 |
What sampling times should be adopted ?This is an operational decision which should be made by the operations manager and the RPA. |
Q8A8 |
Why are particular flow rates used?To keep a reasonable pressure differential across a small filter and also to limit power usage. In any case the rates used are comparable to human breathing rates (which are typically 37 litres per minute). |
Q9A9 |
What are the effects of long-term deposition on filter paper, and how should the filters be maintained?Accumulation of dust may result in:
In the absence of an instrument differential pressure-drop monitor function, filters should be changed according to ambient dust levels (e.g. during facility demolition this may need to be done hourly, but in clean rooms maybe once a week would be sufficient). This decision lies with the local Responsible Person, taking advice from the RPA. |
Q10A10 |
What are the influences of pipe fittings between the sampling point and detector?Generally, pipes should be kept as short as possible. The influence of pipe diameter is that narrow pipes result in a large pressure drop and a high flow rate (resulting in impaction on bends), whereas wide pipes result in a low flow rate and deposition of particulates on horizontal surfaces. So it is advisable to avoid long horizontal pipe runs, narrow pipes and bends. Any bends in the system should have large radii to minimise impaction. |
Q11A11 |
What methods are available for radon compensation?A range of a-particle energies arise from radon isotopes. One simple spectrometric method is to set a wide window around each radon daughter peak and then correct for spillage of signal from high-energy peaks to lower-energy peaks. The method needs to be tailored to the local area. Multi-channel spectroscopy tends to enable better factors and requires less adjustment for the local area. If it is possible to wait for the radon daughters to decay away, this should be done. |
Q12A12 |
Which reference sources should be used for instrument calibration?This subject is covered in detail in the Appendix to NPL Measurement Good Practice Guide No. 82: "The examination and testing of equipment for monitoring airborne radioactive particulate in the workplace".1 Key points are:
|
Q13A13 |
What are the main sources of measurement uncertainty?These include:
NPL can provide advice on the estimation of measurement uncertainties. |
Q14A14 |
Where can Type Test data be found?Manufacturers, suppliers and other organisations may provide these. They are usually based on recommendations from, for example, IEC.2 Some have been adopted as British and European standards.3,4 |
Q15A15 |
Which standards and specifications must be complied with?Ultimately, it is Regulation 19 ('Monitoring of designated areas') of The Ionising Radiations Regulations 19995 which must be complied with. Documents such as NPL Measurement Good Practice Guides1,6-8 are useful in providing guidance in how to meet the requirements of this regulation. |
Q16A16 |
What are other possible sources of information?These include:
|
Q17A17 |
Is there a glossary of terms (e.g. DAC, DAC.h) available?Yes – there is a glossary in NPL Measurement Good Practice Guide No. 82: '‘The examination and testing of equipment for monitoring airborne radioactive particulate in the workplace'.1 |
References:
|
|
Peter Burgess, NPL
Julian Dean, NPL
John Simpson, Consultant
March 2008
For further information, please contact Julian Dean
