Photon beam calibration service for radiotherapy (air kerma and absorbed dose)
Traceable calibration of a therapy-level dosemeter in terms of absorbed dose to water or air kerma. The service is available for thimble chambers or thin-window chambers, and offered in the Spring or Autumn (subject to demand) of each year.
Thimble Chambers
Air kerma (low + medium energy X rays + 60Co) and absorbed dose
Service delivery
- For instruments intended to be used as secondary standards or others required for measurements of the greatest accuracy
- Calibration of an ionisation chamber with or without an electrometer
- Ionisation chambers of type 2561, 2611, 2571, and other Farmer type chambers with a graphite wall and aluminum central electrode will be accepted after confirmation with NPL
- Batch calibrations are undertaken between February and April each year
Absorbed dose to water
- Calibration against NPL reference standards in terms of absorbed dose to water
- Primary standard is the NPL graphite microcalorimeter
- For 60Co γ-rays and X-rays generated between 4 MV and 25 MV, quality index (TPR) between 0.568 and 0.800
- Beam qualities similar to those used for high energy photon radiotherapy.
- Calibration conforms to IPSM Code of Practice (1990)
Air kerma
- Calibration directly against the NPL air kerma primary standards
- Calibration for X-rays generated between 50 kV and 280 kV and for 60Co γ-rays
- Beam qualities similar to those used for radiotherapy treatments
- Calibration with beam qualities selected from Table 1
- Calibration conforms to IPEMB Code of Practice (1996)
Thin-Window Chambers
Air kerma - very low energy X-rays
Service delivery
- For instruments intended to be used as secondary standards or others required for measurements of the greatest accuracy.
- PTW 23342, 0.02 cc, PTW 23344, 0.2 cc and other similar chambers are accepted after confirmation from NPL.
- Calibration of an ionisation chamber with or without an electrometer.
- Batch calibrations are undertaken between February and April each year.
Air kerma
- Calibration directly against the NPL 50 kV air kerma primary standar.
- Calibration for X-rays generated between 8.5 kV and 50 kV
- Beam qualities similar to those used for radiotherapy treatments (see Table 2)
- Calibration with beam qualities selected from Table 2
- Calibration conforms to IPEMB Code of Practice (1996)
Contact
Customer Services tel: +44 20 8943 8695
E-mail: dosimetry_enquiries@npl.co.uk