HDR and LDR brachytherapy
Calibration Service for 192Ir High Dose Rate (HDR) Brachytherapy Dosemeters
NPL's 192Ir HDR brachytherapy calibration service is for ionisation chambers used as secondary standards or instruments required for measurement of the greatest accuracy. Calibration of either an ionisation chamber alone or of a complete system including an electrometer is available.
Service Features
- Calibration directly against the NPL air kerma primary standard
- Reference air kerma rate (RAKR) of the 192Ir source up to 50 mGy/h
Specification and Availability
- Calibration of well-type (re-entrant) chambers suitable for HDR sources
- Calibration of thimble chambers with customer provided calibration jig
- Electrometer calibration (charge or current calibration)
- Calibrations are normally carried out in spring, and are undertaken using a microSelectron 192Ir HDR source
- Calibration conforms to IPEM Code of Practice (2010)
For more information, please contact Thorsten Sander or our Radiation Dosimetry Team
Calibration Service for Low Dose Rate (LDR) Brachytherapy Sources
NPL's LDR brachytherapy service includes the calibration of 125I seeds and 192Ir wires and pins. The sources are calibrated with the Fidelis secondary standard radionuclide calibrator.
Service Features
- Source calibration traceable to the NPL air kerma primary standard
Specification and availability
- Source calibrations in terms of μGy/h at 1 m or MBq
- Calibration of 125I seeds and 192Ir wires and pins, including 125I seeds (e.g. IMC 6711 OncoSeedTM, IMC 7000 RAPIDStrandTM)and 192Ir wires and pins (e.g. Amersham ICW-series)
- Calibrations of other source types/configurations possible
- Calibrations are available by arrangement throughout the year