National Physical Laboratory

Best Practice Guide to Measurement of Acoustic Output Power (Measurement)

IntroductionRadiation Force Balances | Measurement | Additional Information | Further Reading

Uncertainties

Random (Type A evaluation)

The random component of measurement is a good indicator of the stability of a system. Unlike the systematic component, which generally remains unchanged for a standard measurement protocol, the random component can vary depending on the stability of the system and the conditions under which the unit is measured. Generally, a unit which is less stable will show higher random uncertainty.

The method incorporated in an NPL calibration can be found in the UKAS document 'The expression of Uncertainty and confidence in Measurement' and two worked examples of the method have been provided. The first set of data is from a Reference source used as a transfer standard for calibrations at NPL and has itself been calibrated prevously against the NPL Primary Standard Radiation Force Balance. The second set of data is from a second standard off the shelf device.

Data set 1 Data set 2
99.48mW
99.15mW
99.20mW
99.07mW
99.20mW
98.74mW
99.12mW
98.83mW
99.06mW
98.81mW
98.64mW
98.70mW
94.22 mW
96.91mW
95.64mW
97.99mW
96.38mW
98.16mW
96.89mW
98.49mW
97.21mW
99.17mW
97.76mW
99.84mW
Mean value
AOP - Mean
x = 99.00 mW x = 97.39 mW
Standard deviation
AOP - Standard deviation
s(xj) = 0.254 mW s(xj) =1.541 mW
Standard uncertainty
AOP - Standard uncertainty
U(x) = 0.073 mW (0.07%) U(x) = 0.445 mW (0.46%)

Data set two shows an increase in the measured force with an increase in the number of insonation periods. A systematic difference can be seen between the two transitions (OFF-ON and ON-OFF). The variation for the results is reflected in the overall calculated value for the random uncertainty. Calculating each transition can be a good indicator of any performance characteristics of a device and can contribute to an appropriate procedure for measurement. An example for data set 2 would be to use only one set of transitions, say ON to OFF, and a specified insonation period as the basis for deriving a reference value. The method used should be replicated for most accurate transfer to a device under test.

References

Next Page >>

Last Updated: 16 Sep 2010
Created: 16 Sep 2010